Detailed Data Setting; Current Output Setting; Oxygen Analyzer_Current Output Setting; Oxygen Analyzer_Analog Output Setting - YOKOGAWA ZR402G User Manual

Integrated type zirconia oxygen/humidity
Hide thumbs Also See for ZR402G:
Table of Contents

Advertisement

8.

Detailed Data Setting

8.1

Current Output Setting

8.1.1

Oxygen Analyzer_Current Output Setting

This section describes setting of the analog output range. Table 8.1 shows parameter codes for
the set items.
Table 8.1
Set item
Analog output
Output mode
Min. oxygen concentration
Max. oxygen concentration
Output damping constant
*1:
For the oxygen analyzer, set 0 (zero) only for parameter code C01. When it is set, the current output is 4-mA fixed regardless of
the oxygen concentration.
8.1.2

Oxygen Analyzer_Analog Output Setting

This section describes how to set the analog output range.
(1) To provide an oxygen concentration, use parameter code C11 to set the minimum
oxygen concentration at 4 mA, and use parameter code C12 to set the maximum oxygen

concentration at 20 mA.

(2) To provide a humidity output, use parameter code C13 to set the minimum humidity at 4 mA,
and use parameter code C14 to set the maximum humidity at 20 mA.
(3) To provide a mix ratio, use parameter code C15 to set the minimum mix ratio at 4 mA, and
use parameter code C14 to set the maximum mixing ratio at 20 mA.
Refer to Table 8.2 for the parameter codes.
Table 8.2
Set item
Analog output
Output mode
Min. oxygen concentration
Max. oxygen concentration
Min. humidity
Max. humidity
Min. mixing ratio
Max. mixing ratio
Output damping constant
Current Output Parameter Codes
Parameter code
C01
C03
C11
C12
C30
Current Output Parameter Codes
Parameter code
C01
C03
C11
C12
C13
C14
C15
C16
C30
<8. Detailed Data Setting>
Set value
0 Oxygen concentration
1 4 mA (fixed *1)
2 4 mA (fixed *1)
0 Linear
1 Logarithm
Oxygen concentration at 4 mA
Oxygen concentration at 20 mA
0 to 255 seconds
Set value
0
Oxygen concentration
1
Humidity
2
Mixing ratio
0
Linear
1
Logarithm
Oxygen concentration reading corresponding to 4 mA
Oxygen concentration reading corresponding to 20 mA
Humidity reading corresponding to 4 mA
Humidity reading corresponding to 20 mA
Mixing ratio at 4 mA
Mixing ratio at 20 mA
0 to 255 seconds
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
8-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22aZr202gExaxt zr seriesZr202aZr22g

Table of Contents