In a semi-automatic calibration, "during calibration" is the time, starting when a calibration
instruction is executed with an infrared switch or a contact input, to make a series of
calibrations, until the preset output stabilization time elapses.
In an automatic calibration, "during calibration" is the time, starting when automatic
calibration is carried out at the calibration start time, until the preset output stabilization time
elapses.
(4) "Error" appears when Error 1 to Error 4 are being issued
8.2.2

Preference Order of Output Hold Value

The output hold value takes the following preference order:
Preference order (high)
For example, if the output current is set to "4 mA" during maintenance, and "without hold" output
during calibration is preset, the output is held at 4 mA in the maintenance display. However, the
output hold is released at the time of starting the calibration, and the output will be held again at 4
mA after completing the calibration and when the output stabilization time elapses.
8.2.3

Output Hold Setting

Table 8.5 lists parameter codes with set values for individual set items.
Table 8.5
Set items
During warm-up
During maintenance
During calibration
During error occurrence
Note: "C07" is not displayed when option code "/C2" or "/C3" (NAMUR NE 43 compliant) is specified.
8.2.4

Default Values

When the analyzer is delivered, or if data are initialized, output holding is by default as shown in
Table 8.6.
Table 8.6
Status
During warm-up
Under maintenance
Under calibration or blow-back
On Error occurrence
During error occurrence
During calibration
During maintenance
During warm-up
Parameter Codes for Output Holding
Parameter code
C04
C05
C06
C07
Output Hold Default Values
Output hold (min. and max. values)
4 mA
Holds output at value just before maintenance started.
Holds output at value just before starting calibration
Holds output at a preset value.
<8. Detailed Data Setting>
8.3.2E.siki
Set value
0 4 mA
1 20 mA
2 Holds Set value
0 Without hold feature
1 Last measured value.
2 Holds set values.
0 Without hold feature
1 Last measured value.
2 Holds set values.
0 Without hold feature
1 Last measured value.
2 Holds set values.
Preset value
4 mA
4 mA
4 mA
3.4 mA
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00