3.2 Setting the Output Conditions in the DC Mode

Selecting the Output Channel
Setting the Output Voltage Range
Setting the Output Voltage
Setting the Output Mode
3-2
In this section, you will be using the operation panel that appears by double-clicking the
module icon in the station list window as shown in the figure below and the trigger
operation panel that appears by clicking the Trig tab.
Trig tab
Select the output channel
Set the output voltage range
Check the [On] box of the channel you wish to output.
Select the output voltage range from [1 V], [2 V], [5 V], and [10 V] in the [Range] list box.
Set the output voltage in the [Output] entry box. The output voltage can be set in units of
0.0001 V in the 1 V range and 0.001 V for all other ranges.
Select the output mode using the [Output Mode] option button. The choices are shown
below. [Cont] is default.
Cont: Continuous output
Continuously outputs DC voltage from the time the [Output] button is clicked to turn the
output ON until the next time it is clicked to turn the output OFF.
Trigger: Trigger output
After the [Output] button is clicked to turn ON the output, DC voltage is continuously
output when a trigger occurs. If the output voltage is changed while the waveform is
being output, the new DC voltage is output the next time the trigger occurs.
Gate: Gate output
After the [Output] button is clicked to turn ON the output, the output starts when the gate
is enabled and stops when it is disabled. The output voltage can be changed while the
gate is enabled.
Note
When the output is OFF, the waveform output terminal of the module outputs 0 V.
Set the output voltage
Select the output mode
Execute manual trigger
Turn ON/OFF
waveform output
IM 707281-01E