Stop valve
or
Check valve
Reference gas
Calibration gas
Figure 4.3
System 2 illustrated in Figure 4.3 requires piping as follows:
• Mount the check valve or the stop valve through a nipple to the calibration gas inlet of the
equipment.
4.2.1

Piping Parts for System 2

Check that the parts listed in Table 4.2 are provided.
Table 4.2
Equipment
Oxygen/
Calibration gas inlet Stop valve or check
Humidity
Analyzer
Reference gas inlet
Note: Parts with marking * are used when required.
General parts can be found on the local market.
4.2.2

Piping for the Calibration Gas

This piping is to be installed between the zero gas cylinder and the ZA8F flow setting unit, and
between the ZA8F flow setting unit and the ZR202G analyzer.
The cylinder should be placed in a case assembly E7044KF or the like to avoid any direct
sunlight or radiant heat so that the gas cylinder temperature may not exceed 40°C.
Mount the pressure regulator (recommended by YOKOGAWA) on the cylinder.
Mount the stop valve or the check valve (recommended by YOKOGAWA) through the nipple
(found on the local market) at the calibration gas inlet of the equipment as illustrated in Figure
4.4. (The stop valve or the check valve may have been mounted on the equipment when
shipped.) Connect the flow setting unit and the analyzer to a 6mm (O.D.) x 4mm (I.D.) (or nominal
size 1/4 inches) or larger stainless steel pipe.
ZR202G Integrated type Zirconia Oxygen/Humidity Analyzer
~
100 to 240 V AC
Contact input
Analog output, contact output
Digital output (HART)
Needle
Flowmeter
valve
ZA8F flow setting unit
Piping for System 2
Piping Parts
Piping location
Parts Name
valve
Nipple *
Zero gas cylinder
Pressure Regulator
Joint for tube connection Rc1/4 or 1/4 NPT
Air set
Joint for tube connection Rc1/4 or 1/4 NPT
Air Set
Instrument air
Span gas(Same as Zero gas calibration)
Pressure
regulator
Zero gas cylinder
Calibration gas unit case
F1.2E.ai
Stop valve (L9852CB or G7016XH
Check valve (K9292DN or K9292DS)
Rc1/4 or 1/4 NPT
User' s scope
(G7013XF or G7014XF) recommended by
YOKOGAWA
(G7003XF/K9473XK or G7004XF/
K9473XG) recommended by YOKOGAWA
<4. Piping>
Description
recommended by YOKOGAWA
provided by YOKOGAWA
General parts
General parts
General parts
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
4-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr22aZr202gExaxt zr seriesZr202aZr22g

Table of Contents