Table 8.14
Switch operation
>
>
>
>
>
>
>
>
>
>
>
>
The symbol (
8.6.2
Setting Periods over which Average Values are
Calculated and Periods over which Maximum and
Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maximum and
minimum values under measurement (see Section 10.1, later in this manual). The following
section describes how to set the periods over which oxygen concentration average values are
calculated and maximum and minimum values are monitored.
Procedure
n
Use the parameter-code table below to set the average, maximum and minimum oxygen
concentration values. Periods over which average is calculated and periods over which
maximum and minimum values are monitored can be set, ranging from 1 to 255 hours. If the set
ranges are beyond the limits specified, an "Err" will be displayed.
Data-and-time Settings
Display
F10
ENT
00.01.01
ENT
00.01.01
ENT
00.06.01
ENT
00.06.01
ENT
00.06.21
ENT
00.06.21
ENT
07.18
ENT
14.30
ENT
14.30
ENT
14.30
ENT
F10
ENT
) indicates that the corresponding keys are being touched, and the light characters indicate flashing.
<8. Detailed Data Setting>
Description
Select the parameter code F10.
If you touch the [ENT] key, the current date will be displayed. The
display on the left indicates the date - January 1, 2000. To set
June 21, 2000, follow the steps below:
Touch the [>] key to move the position of the digit that is flashing
to the right.
Touch the [
] key to change to 6.
Touch the [>] key to move the position of the digit that is flashing
to the right one digit.
Touch the [
] key to change to 2.
Touch the [>] key to move the position of the digit that is flashing
to the right one digit.
Let the rightmost character flash, and touch the [>] key to display
the time.
Continuously touch the [>] key, then the date and time are
alternately displayed. Displayed on the left is 7:18 a.m.
Omitted here.
Touch the [
] key and enter the current time in same way as the
date has been entered, on a 24-hour basis. 2:30 p.m.
Displayed on the left means 2:40 p.m.
If you touch the [ENT] key, all the digits flash.
Touch the [ENT] key again to set the time.
If you touch the [>] and [ENT] keys together, the parameter code
selection display appears.
IM 11M12A01-04E