Duct Adapter For Remote Sensor - Dräger Polytron 7000 Instructions For Use Manual

Approved as type p3u and p3fb. transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Connecting to the Polytron 7000
Remote sensor adapter
4 Connect the plug of the Remote Cable (cable length 5, 15 or 30 m) to the
remote sensor adapter and secure it by turning the ring clockwise.
5 Insert the sensor plug in the opening on the Dräger Polytron 7000 transmit-
ter with the "Dräger" logo pointing to the front.
Secure the sensor plug with the bayonet ring.
The remote sensor is recognised automatically by the transmitter. No further
installation steps are required.

Duct adapter for remote sensor

Purpose
Duct adapter for remote sensor – 83 17 617:
— For mounting a remote sensor on a pipe or duct,
— For measuring the gas concentration in the pipe or duct.
Mounting
Drill a hole for the sensor opening (diameter 35+1 mm) at the desired meas-
uring point on the pipe.
Button the sealing sleeve into the hole.
Align the retaining clip so that it is centred on the hole.
1 Drill the holes for the securing screws. A drilling template is provided on
page 141.
2 Loosen the bayonet ring of the remote sensor.
3 Place the retaining clip on the pipe of the case and install the bayonet ring
again.
Insert the sensor opening into the sealing sleeve.
Turn the retaining clip to the correct position and screw it down.
4
Connector
NOTE
Remote Cable
5 m / 15 m / 30 m
3
Installing accessories
5
Sensor plug
1
1
2
29