Dräger Polytron 7000 Instructions For Use Manual page 49

Approved as type p3u and p3fb. transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Fault number
Cause
# 130
The function » Sensor lock « is active. A
sensor with a different Part No. has
been inserted.
# 134
Bad contact between the sensor and
the sensor card.
# 136
Sensor hardware fault.
Warning number Cause
# 1
Data error in the unit. Certain functions
such as the Datalogger or the dongle
functions may not be available.
# 51
Datalogger is in stack mode and is 100
% full. No more data can be recorded.
# 52
Datalogger is in stack mode and is 90
% full.
# 53
No valid date and/or time is set.
# 58
Software dongle was removed without
logging off.
Hardware fault in the software dongle.
# 64
Pump gas flow falls below the warning
threshold.Reliable measurements are
no longer possible.
# 106
Increased zero-point offset.
# 111
Sensor is not working in the specified
temperature range.
# 112
Sensor near end of life.
# 114
Calibration interval expired.
# 115
Sensor was operate for too long with a
high concentration.
# 119
Sensor is not yet fully warmed up. An
increased measuring error must be
expected.
# 120
The sensor has been exposed to an
excessively high gas concentration for
a long period.
# 131
Bad sensor, no longer operating.
# 132
Electrolyte liquid evaporates. Measure-
ments will soon no longer possible.
# 135
Information such as the Part No. and
the Serial No. is not available.
Remedy
Deactivate the function » Sensor lock «, page 77 or use
a sensor with the same Part No. as the one which
was removed.
Check the sensor contacts. Remove and reinstall the
sensor several times. If the problem persists, install a
new sensor, page 46.
Remove the sensor and install it again. If the problem
persists, install a new sensor, page 46.
Remedy
Initialise the unit with the menu items » Settings «,
» Instrument «, » Init. device «, page 68.
If this error occurs again: have the transmitter checked
by DrägerService.
Read out the data. Then clear and restart the Datalog-
ger
Read out the data as soon as possible. Then clear and
restart the Datalogger.
Set the date and time, page 67.
Deactivate the function in the menu » Settings «,
» Instrument «, » SW dongle «, » XXX dongle «,
page 69.
Change the software dongle, page 30.
Check the hoses for blockages,
If necessary adjust pump flow.
Zero-point calibration, page 58.
Operate sensor in the specified temperature range (see
Instructions for Use of the sensor).
Replace the sensor, page 46.
Recalibrate the unit, page 44.
Reduce overgassing.
Wait until the sensor has fully warmed up.
Reduce the exposure to high gas concentrations. If this
does not help, replace the sensor, page 46.
Replace the sensor, page 46.
Refill electrolyte. See sensor data sheet.
Disconnect the unit from the mains and restart it.
If this error occurs again: have the transmitter checked
by DrägerService.
Fault – Cause – Remedy
49