Duct Adapter For Remote Sensor - Dräger Polytron 7000 Instructions For Use Manual

Transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Installing accessories
Connecting to the Polytron 7000
Remote sensor adapter
4 Connect the plug of the Remote Cable (cable length 5, 15 or 30 m) to the remote
sensor adapter and secure it by turning the ring clockwise.
5 Insert the sensor plug in the opening on the Dräger Polytron 7000 transmitter
with the "Dräger" logo pointing to the front.
Secure the sensor plug with the bayonet ring.
Attention:
For use in Zone 22, tighten the locking screw (2 mm Allen screw) of the
sensor bayonet ring tight enough to ensure that the bayonet ring is
secured against unintended loosening.
Note:
The remote sensor is recognised automatically by the transmitter. No further
installation steps are required.

Duct adapter for remote sensor

Purpose
Duct adapter for remote sensor – 83 17 617:
— For mounting a remote sensor on a pipe or duct,
— For measuring the gas concentration in the pipe or duct.
Mounting
Drill a hole for the sensor opening (diameter 35+1 mm) at the desired measuring
point on the pipe.
Button the sealing sleeve into the hole.
Align the retaining clip so that it is centred on the hole.
1 Drill the holes for the securing screws. A drilling template is provided on
page 137.
28
Remote Cable
4
5 m / 15 m / 30 m
Connector
5
Sensor plug
1
1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

P3uP3fb