Installing Accessories; Daisy Chain Kit - Cable Entry Kit - Dräger Polytron 7000 Instructions For Use Manual

Approved as type p3u and p3fb. transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Installing accessories

Various accessories are available for the Dräger Polytron 7000 transmitter
and may also be installed later.
Daisy chain kit – Cable Entry Kit
Intended use
Daisy chain kit – 83 17 282:
— For the connection of several transmitters to one bus cable (daisy chain or
multidrop connection)
Contents of the kit
— Cable gland with nut, 4-pole orange terminal block
Preparing the docking station
Install the docking station as described in the installation notes 90 23 760.
1 Break or drill out the prepared breakthrough for the second cable gland
from the inside of the docking station.
The hole should have a diameter of 20.5 mm.
Check that the docking station has no loose parts, and clean it if necessary.
Insert the nut of the cable gland into the docking station.
Screw in the cable gland from the outside of the docking station and tighten
it.
Electrical connections
Connect the wiring as described in the installation notes for the docking
station (90 23 760).
2 Insert the 4-pole terminal block into the holder on the mounting plate.
If necessary:
refit the raincover (protection against dust and splashing water).
The two 4-pole terminal blocks for input (3) and output (2) are electrically
connected 1:1 to each other inside the transmitter.
Installing the measuring unit
Install the measuring unit as described in the installation notes 90 23 759.
— Due to the second terminal block, the resistance when installing the meas-
uring unit is slightly higher.
Make sure that the measuring unit is fully lowered and hits the stop.
NOTE
Installing accessories
1
X7
X8
2
3
27

Hide quick links:

Advertisement

Table of Contents
loading