Electrical connections
Installing fieldbus communication on the transmitter
— The PROFIBUS PA transmission technology for intrinsically safe applica-
tions is MBP. MBP stands for:
– Manchester Coding (M)
– Bus Powered (BP)
— 5 transmitters can be connected to a segment with a typical segment cur-
rent of 100 mA.
Install the 4-wire connection cable in the cable gland, cut it to length and
strip off the insulation (approx. 80 mm).
Shorten the shield (if installed) to prevent short-circuiting.
Connect cable:
1 4-pin terminal block for Dräger Polytron 7000, observing the polarity.
2 4-pin terminal block for Dräger Polytron 7000, observing the polarity.
Slide connecting terminal back into holder.
Secure cable in holder.
Fold up the installation notes and place them in the Dräger docking station
for future use during commissioning.
Refit raincover (protection against dust and splashing water).
Installing the transmitter in 4-wire connection:
With 4-core connection the transmitter has no Ex protection. Once the trans-
mitter has been used after installation in this manner, it may never be
installed in explosion-hazard areas. Explosion hazard!
24
CAUTION
X7
X8
1
2