Submenu » Sensor - Dräger Polytron 7000 Instructions For Use Manual

Approved as type p3u and p3fb. transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

The menu » Settings «
Use the »
« and »
to activate this selection.
— The Neuron ID of the Polytron 7000 is transmitted.
PA/FF interface
Provides for the address distribution for the Profibus PA interface.
Select » Settings «, » Communication «, » PA/FF Interface « and » PA
address « in this order, pressing the »
Select the line for the address input by using the »
with the »
« key.
OK
Address entry by pressing the »
of the input position with subsequent address entry (range 1-126) by press-
ing the »
« or »
M
Confirm adjusted address with the »
Submenu » Sensor «
The submenu » Sensor « can be used to make various settings for the
installed sensor.
Change sensor
— This function can be used to change a sensor while the unit is running with-
out sending a fault alarm to the control unit. It also ensures that all sensor
data currently in the microprocessor can be saved to the EEPROM of the
sensor before the sensor plug is disconnected.
— In principle, a sensor can be replaced at any time. However, for technical
safety reasons, a fault alarm will be activated until a new sensor is connect-
ed, in case a sensor is disconnected accidentally.
Select the menu items » Settings «, » Sensor « and » Change sensor « in
this order, pressing the »
— The maintenance signal is transmitted on the analogue interface, and the
display reads: » Please remove sensor«.
Remove the old sensor and plug in the new one (see page 46).
When the new sensor has been installed:
The message » Loading database, please wait « is displayed.
When the sensor data have been loaded:
— The message » Database is loaded « is displayed.
Select » Back to menu « and press the »
— The maintenance signal on the analogue interface remains until the sensor
operates normally.
The duration of the warming-up period depends on the type of sensor and
its history. If, for example, the sensor had already been warmed up on an-
other transmitter and it was disconnected only for a short time, the warm-
ing-up period may be shorter than that shown in the operating instructions
for the sensor.
See the operating instructions for the sensor for the maximum warming-up
period.
If the old sensor is replaced with an identical sensor (with the same Part
No.), the configuration of the transmitters (gas type, measuring range, cal-
ibration gas, calibration interval) remains unchanged.
Otherwise, the default values for the sensor (see the operating instructions
76
« to select » Confirm « and press the »
M
« key after each selection.
OK
« key for selecting the desired position
OK
« keys.
« key.
OK
« key after each selection.
OK
« key.
OK
« key
OK
« key and confirm
M