Pump module
The pump module is not covered by the explosion protection approvals.
Use is not permitted in explosion-hazard areas! Explosion hazard!
Supply voltage (DC)
— for an output of 0.5 L/min
— for an output of 1.0 L/min
— for an output of 1.5 L/min
Power consumption
— for an output of 0.5 L/min
— for an output of 1.0 L/min
— for an output of 1.5 L/min
Output
— Setting range
— Factory setting
Flow warning
— Factory setting
Flow alarm
— Factory setting
Hose connectors
— Internal hose diameter
Materials used in the gas path
— Pump adapter
— Gas guides
— Dust filter
— Pump
Ambient conditions
for operation
during storage
CE markings
CAUTION
12 V to 30 V at the transmitter
16 V to 30 V at the transmitter
20 V to 30 V at the transmitter
<2 W
<4 W
<6 W
approx. 0.5 l/min to 1.5 l/min (approx. 30 % to 100 %)
0.5 L/min
0.4 L/min
0.3 L/min
5 mm
PP, Viton
PP, glass
PE
EPDM, PTFE, Niro
o
o
0 to 55
C (32 to 130
F)
700 to 1300 hPa (20.7 to 38.4 inch Hg)
0 to 100 % relative humidity, non condensing
o
–40 to 70
C (–40 to 150
700 to 1300 hPa (20.7 to 38.4 inch Hg)
0 to 100 % relative humidity, non condensing
— Electromagnetic compatibility (Directive 2004/108/EC), when used with
transmitter
o
F)
Technical Data
91