Electrical connections
Transmitter with pump module (3-wire):
Minimum
Maximum pos-
supply voltage
sible
flow-rate
20 V
0.5 L/min
1.0 L/min
24 V
1.5 L/min
28 V
Transmitter with relay and pump modules (3-wire):
Minimum
Maximum pos-
supply voltage
sible
flow-rate
20 V
0.5 L/min
1.0 L/min
24 V
1.5 L/min
28 V
22
Maximum cable length with a load resistance of not more than
Number of
transmitters
0.5 mm
265 pF/m
1
161 m
2
80 m
3
54 m
4
40 m
1
233 m
2
116 m
3
78 m
4
58 m
1
135 m
2
68 m
3
45 m
4
34 m
1
271 m
2
135 m
3
90 m
4
68 m
Maximum cable length with a load resistance of not more than
Number of
transmitters
0.5 mm
265 pF/m
1
110 m
2
55 m
3
37 m
4
27 m
1
161 m
2
80 m
3
54 m
4
40 m
1
90 m
2
45 m
3
30 m
4
22 m
1
179 m
2
90 m
3
60 m
4
45 m
500 Ω
2
2
0.75 mm
320 pF/m
241 m
120 m
80 m
60 m
349 m
175 m
116 m
87 m
203 m
102 m
68 m
51 m
406 m
203 m
135 m
102 m
500 Ω
2
2
0.75 mm
320 pF/m
164 m
82 m
55 m
41 m
241 m
120 m
80 m
60 m
134 m
67 m
45 m
34 m
269 m
134 m
90 m
67 m
2
1.5 mm
2.5 mm
375 pF/m
400 pF/m
482 m
792 m
241 m
401 m
161 m
268 m
120 m
201 m
699 m
792 m
349 m
582 m
233 m
388 m
175 m
291 m
406 m
586 m
203 m
338 m
135 m
226 m
102 m
169 m
812 m
792 m
406 m
677 m
271 m
451 m
203 m
338 m
2
1.5 mm
2.5 mm
375 pF/m
400 pF/m
329 m
548 m
164 m
274 m
110 m
183 m
82 m
137 m
482 m
792 m
241 m
401 m
161 m
268 m
120 m
201 m
269 m
448 m
134 m
224 m
90 m
149 m
67 m
112 m
537 m
792 m
269 m
448 m
179 m
298 m
134 m
224 m
2
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents