Fault – Cause – Remedy
Fault – Cause – Remedy
If the display will not function: Have the transmitter checked by
DrägerService.
The fault and warning numbers shown in the following tables are displayed in
the menu under » Information «, » Instrument «, » Fault « or » Warnings « –
see page 55.
Fault number
Cause
# 1
Serious data error in unit –
various causes.
# 2
Serious unit fault –
various causes.
# 61
Data error on the interface card –
various causes
# 63
Hardware fault or software error in the
pump module.
# 64
Gas flow of the pump module falls
below the configured fault threshold.
Reliable measurements are no longer
possible.
# 65
Open-circuit in 3-wire cable.
# 67
Bad contact of the relay module.
# 100
Unit cannot detect a sensor.
# 101
Sensor data error in the unit.
# 102
Unit does not support this
sensor version.
# 103
Sensor data error in the unit.
# 106
Zero-point not correctly.
# 107
Sensitivity calibration not executed cor-
rectly.
# 108
Sensor data error.
# 109
Unit fault.
# 121
Fresh-air calibration (first step of auto-
calibration) not executed correctly.
# 125
Sensor not ready.
# 129
Electrolyte liquid evaporates
48
Remedy
Initialise the unit with the menu items » Settings «,
» Instrument «, » Init. device «, page 68.
If this error occurs again: have the transmitter checked
by DrägerService.
Have the transmitter checked by DrägerService.
Have the transmitter checked by DrägerService.
Change the pump module, page 34.
Check the hoses for blockages,
If necessary adjust pump flow.
Check the connections.
Check connector of the relay module or fit it again.
Remove the sensor and install it again, page 46.
If the problem persists, check the sensor plug or install
a new sensor.
Remove the sensor and install it again, page 46.
If this error occurs again: have the transmitter checked
by DrägerService.
Use a compatible sensor, see the ordering list on
page 93.
Initialise the sensor with the menu items » Settings «,
» Sensor «, » Sensor-EC «, » Init. sensor «, page 78.
If this error occurs again: have the transmitter checked
by DrägerService.
Zero-point calibration, page 58.
Repeat sensitivity calibration, page 59.
Replace the sensor, page 46.
Check the sensor contacts; otherwise have the transmit-
ter checked by DrägerService.
Repeat autocalibration, page 60. Make sure that the
ambient air is free of other gases.
Replace the sensor, page 46.
Refill electrolyte. See sensor data sheet.