Dräger Polytron 7000 Instructions For Use Manual page 122

Approved as type p3u and p3fb. transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

Metrological certificate of approval
Gas / vapour
ethylene
ethine
ethylene oxide
hydrogen fluoride
formaldehyde
carbon dioxide
carbon monoxide
methyl methacrylate
methylamine
carbonyl chloride
phosphine
i-propyl alcohol
propylene
sulphur dioxide
hydrogen sulfide
hydrogen selenide
hydrosilicon
nitrogen dioxide
nitrogen monoxide
styrene
tetrahydrofurane
tetrahydrothiophene
vinyl chloride
hydrogen
hydrogen peroxide
(–) * Negative display.
The influence of unsaturated hydrocarbons, alcohols or hydrogen in higher concentrations or over a longer period of
time (dose approx. 100,000 ppm x hours) will let the sensor drop out.
When used in transmitters with pump module, the O
locations. If used in this combination, vibrations may cause the measured value to deviate outside of the permissible
range.
122
Chemical symbol
C
H
2
4
C
H
2
2
C
H
O
2
4
HF
HCHO
CO
2
CO
CH
C(CH
)COOCH
2
3
3
CH
NH
3
2
COCl
2
PH
3
(CH
)
CHOH
3
2
CH
CHCH
3
2
SO
2
H
S
2
SeH
2
SiH
4
NO
2
NO
C
H
CHCH
6
5
2
C
H
O
4
8
C
H
S
4
8
C
H
Cl
2
3
H
2
H
O
2
2
CAUTION
LS sensor (68 09 630) must always be installed at vibration-free
2
Gas concentration
Measurement value devia-
2 Vol.-%
1 Vol.-%
20 ppm
15 ppm
40 ppm
5 Vol.-%
100 ppm
50 ppm
100 ppm
1 ppm
10 ppm
500 ppm
50 ppm
20 ppm
20 ppm
5 ppm
5 ppm
50 ppm
20 ppm
30 ppm
60 ppm
5 ppm
50 ppm
1 Vol.-%
5 ppm
tion in Vol.-%
with dust filter
≤ 1 (–) *
≤ 0.5 (–) *
≤ 2 (–) *
no influence
no influence
no influence
≤ 0.1 (–) *
no influence
no influence
no influence
no influence
no influence
≤ 0.2 (–) *
no influence
no influence
no influence
no influence
no influence
no influence
no influence
no influence
no influence
no influence
≤ 1.5 (–) *
no influence

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents