Autocalibration - Dräger Polytron 7000 Instructions For Use Manual

Approved as type p3u and p3fb. transmitter for electrochemical sensors
Hide thumbs Also See for Polytron 7000:
Table of Contents

Advertisement

The menu » Calibration «
If these are correct as displayed:
Select » Next « and press the »
Disconnect the flow of calibration gas.
An alarm could be triggered in the central unit for as long as the calibration
gas concentration is pending!
— The calibration interval and the date of the next calibration are displayed.
Select » Back to menu « and press the »
Sensitivity calibration can be aborted at any time:
Use the »
« key to move to » Back « and press the »

Autocalibration

Autocalibration consists of fresh-air calibration followed by sensitivity calibra-
tion. This function is intended only for users who are familiar with the unit,
since the sequence of actions may result in calibration errors.
The function can be activated or deactivated with the menu sequence
» Settings «, » Sensor «, » Autocal. «.
Calibration procedure (using an EC-H
Select the menu items "Calibration" and "Autocal." in this order, pressing
the »
« key after each selection.
OK
The sensor must be supplied with fresh air, otherwise calibration errors may
occur!
— The maintenance signal is transmitted on the analogue interface.
— The message » Fresh-air calibration running « is displayed.
Expose the sensor to the ambient air (for some sensors, a zero gas such
as N
must be used – see the information in the section "Maintenance" on
2
page 44).
Then:
— The values for the calibration gas are then displayed, for example:
Cal. gas
: H
S
2
Unit
: ppm
Concentr.
: 000025
— The current value and the calibration gas concentration are now displayed.
The calibration operation depends on whether the selected calibration gas
supports so-called autostability.
— Autostability is supported:
When the displayed value has stabilised:
Select » Accept value « and press the »
— Autostability is not supported:
The instrument evaluates the stability of the signal and automatically exe-
cutes the calibration procedure.
60
« key.
OK
CAUTION
« key.
OK
S sensor as an example):
2
CAUTION
« key.
OK
« key.
OK