21
TECHNICAL SPECIFICATIONS
NOTE: SEE TECHNICAL MANUAL FOR USER'S STATEMENT
21.1
Technical data for Planmeca ProMax product family
Classification:
•
Medical Device Directive
•
IEC 60601-1
•
CISPR 11
Generator
X-ray tube
Focal spot size
Total filtration:
•
3D
•
Pan / ceph
Anode voltage
(X-ray units with SW prior to 3.0):
•
3D
•
Pan / SmartPan
•
Ceph
Anode voltage
(X-ray units with SW 3.0 or later):
•
3D
•
Pan / SmartPan
•
Ceph
Anode current
(X-ray units with SW prior to 3.0):
•
3D
•
Pan / SmartPan
User's Manual (3D)
TECHNICAL SPECIFICATIONS
93/42/EEC (Class IIb)
Class I, type B
Class B
Resonant-mode, DSP-controlled, 80...160 kHz,
according to IEC 60601-2-7: 1998
2D / 3D s / 3D Classic / 3D Plus / 3D Mid:
Toshiba D-054SB
3D Max: Toshiba D-067SB
According to IEC 60336
2D / 3D s / 3D Classic / 3D Plus / 3D Mid:
0.5 x 0.5 mm
3D Max: 0.6 x 0.6 mm
min. 2.5 mm Al + 0.5 mm Cu
min. 2.5 mm Al
3D s / 3D Classic / 3D Mid: 54 - 90 kV ±5%
3D Max: 54 - 96 kV ±5%
54 - 84 kV ±5%
60 - 84 kV ±5%
3D s / 3D Classic / 3D Plus / 3D Mid:
60 - 90 kV ±5%
3D Max: 60 - 96 kV ±5%
60 - 84 kV ±5%
60 - 84 kV ±5%
3D s / 3D Classic / 3D Mid: 1 - 14 mA ±10%
3D Max: 1 - 12.5 mA ±10%
1 - 16 mA ±10%
Planmeca ProMax 3D s / 3D 83