3D EXPOSURE
Volume bottom light
Down
56 Planmeca ProMax 3D s / 3D
The following positioning lights will come on:
Guide the patient to the X-ray unit. The patient can sit or
stand during the exposure.
Adjust the X-ray unit to suit the height of the patient. To do
this, press either of the height adjusting buttons until the
chin cup / chin support is at the correct height.
The X-ray unit moves slowly at first, then faster.
Up
Ask the patient to grasp the patient handles.
Check that the patient's head is firmly positioned in the
head support.
If you are using the adjustable head support:
•
You can adjust the head support by turning the
adjusting knob at the top.
•
You can use fastening straps for additional head
support if needed. Refer to section "Head support
A: Attaching adjustable head support" on page 49
for details.
Volume center lights
Incisor light
User's Manual (3D)