18 Technical specifications

18 Technical specifications

18.1 Technical data

Classification
Medical Device Directive
RoHS
IEC 60601‐1
CISPR 11
IP Classification
Applied parts (according to IEC 60601‐1: 2012)
Patient supports
Patient handles
Generator (according to IEC 60601‐2‐7: 1998)
X‐ray tube
Focal spot size (according to IEC 60336: 2005)
Filtration
3D
Pan (SmartPan) / ProCeph
Tube housing front cover quality equivalent
filtration (not included in the specified total
filtration)
Anode voltage
3D
Pan (SmartPan)
ProCeph
Anode current
3D
Pan (SmartPan)
ProCeph
mAs range
Dose range and accuracy
76
Planmeca Viso G7
NOTE
Refer to the Technical Manual for details.
93/42/EEC (Class IIb)
2011/65/EU
Class I, type B
Class B
IPX0
As shown in section Patient supports in User
Manuals
Resonant-mode, DSP-controlled, 80 ‐ 160 kHz
Toshiba D-059SBR or SXR 130-10‐0.5 SC
0.5 x 0.5 mm
Total 2.5 mm Al + 0.2 mm / 0.5 mm Cu
Total 2.5 mm Al
0.3 mm Al @ 70 kV / HVL 2.6 mm Al
80 ‐ 120 kV ±5%
60 ‐ 84 kV ±5%
60 ‐ 84 kV ±5%
Toshiba D-059SBR: 1‐12.5 mA ±10%
SXR 130‐10‐0.5 SC: 1‐16 mA ±10%
Toshiba D-059SBR: 1‐14 mA ±10%
SXR 130‐10‐0.5 SC: 1‐16 mA ±10%
Toshiba D-059SBR: 14 mA ±10%
SXR 130-10‐0.5 SC: 16 mA ±10%
min. / max. as indicated ±(10% + 0.2 mAs)