15

TECHNICAL SPECIFICATIONS

Generator
X-ray tube
Focal spot size
Cone diameter
Max. symmetrical radiation field
Total filtration
Inherent filtration
Anode voltage
Anode current
Target angle
Exposure times
Reference current time product
Lowest current time product
Max. nominal anode voltage
Max. electrical output
Electrical output at 0.1 sec.
Max. loading energy
SSD (Source-Skin Distance) Standard/Long
Long with rectangular collimator
Mains voltage
Mains frequency
Fusing
Duty cycle
Electrical classification
Weight
Ambient temperature
Humidity
User's manual
TECHNICAL SPECIFICATIONS
Constant potential, microprocessor
controlled,
operating frequency 66 kHz
Toshiba D-0711SB
0.7 mm
according to IEC 336
60 mm (2.36 in.)
Rectangular 35 x 45 mm (1.38 x 1.77 in.)
ø 60 mm at SSD 200 mm
ø 60 mm at SSD 300 mm
according to IEC 806
min. 2 mm Al equivalent at 70 kV
according to IEC 522
1 mm Al equivalent at 70 kV
according to IEC 522
50, 52, 55, 57, 60, 63, 66, 70 kV, ±2 kV
8 mA
16°
0.01-3.20 sec., 23 steps
8 mAs at 70 kV, 8 mA, 1 sec.
0.08 mAs at 8 mA, 0.01 sec.
70 kV
560 W at 70 kV, 8 mA
560 W at 70 kV, 8 mA
1800 mAs/h at 70 kV
200 mm (8 in.)/300 mm (12 in.)
306 mm (12.04 in.)
100 V~/110-115 V~/220-240 V~ ±10%
50/60 Hz ±10%
8 AT at 220-240V~
15 AT at 100 V~/110-115 V~
1:15, automatic control
Class I, type B
total 23 kg (51 lbs)
tube head 4.5 kg (10 lbs)
operating +5°C - +40°C
storage -10°C - +50°C
25% - 75%
Prostyle Intra X-ray Unit 27