Planmeca ProMax 3D S User Manual page 76

Dental x-ray unit, 3d imaging
Hide thumbs Also See for ProMax 3D S:
Table of Contents

Advertisement

3D MODEL EXPOSURE
(Part number 10030330)
Adjustable
adapter
in highest
position
Patient support table
72 Planmeca ProMax 3D s / 3D
Polystyrene disc
NOTE The required exposure values depend on the material
Exposure values for 3D Model programs
MATERIAL DENSITY
1
2
3
4
5
Remove any patient supports attached to the X-ray unit
patient support table. Insert the polystyrene disc provided
(part number 10030330) into the adjustable adapter.
Position the adjustable adapter in the highest position as
described in section 11.3 "Adjusting adapter height" on
page 51.
On the X-ray unit, select:
3D Model > 3D Impression scan for impression
material or
3D Model > 3D Plaster cast scan for plaster material.
Material density setting 3.
The exposure values will automatically change to
correspond with the selected material density. The preset
exposure values for each material density are shown in
the following table.
and X-ray unit (X-ray tube and sensor) used. You may
need to select a different material density setting (1, 2,
4 or 5).
kV VALUE
mA VALUE
70
75
80
85
90
14
14
14
14
14