Planmeca ProMax 3D S User Manual page 9

Dental x-ray unit, 3d imaging
Hide thumbs Also See for ProMax 3D S:
Table of Contents

Advertisement

CAUTION
CLASS 1 LASER PRODUCT
APPAREIL À LASER DE CLASSE 1
IEC 60825-1:2007
User's Manual (3D)
NOTE Before taking an exposure, ask any female patient of
NOTE FOR CANADIAN USERS:
NOTE If the X-ray unit has been stored at temperatures under
NOTE Ensure efficient air conditioning in the X-ray room. It is
NOTE If exposures are taken in rapid succession the X-ray
NOTE If the X-ray system is not connected to an
LASER RADIATION -
DO NOT STARE
NOTE FOR US & CANADIAN USERS:
INTO BEAM
1mW
CLASS II
635nm
LASER
PRODUCT
NOTE FOR EUROPEAN USERS:
NOTE EMC requirements have to be considered, and the
NOTE Portable and mobile RF communications equipment
NOTE External equipment intended for connection to signal
NOTE Contact your service technician if you notice a
NOTE Contact your service technician if you have taken an
NOTE Never place or hang any objects on any part of the X-
childbearing age whether she might be pregnant. The
X-ray unit is not intended for use on pregnant women.
All patients must be provided with a shielded apron for
gonad protection and a thyroid shield. The use of a
thyroid shield is especially important in children. The
shielded apron and thyroid shield should have a lead
equivalence of at least 0.25 mm on both sides (front
and back of the patient).
+10°C for more than a few hours, time must be allowed
for the unit to reach room temperature before turning
it on.
recommended to keep the room temperature between
+20°C and +25°C at all times.
tube may overheat and a cooling time will flash on the
touch screen. The cooling time indicates the delay
before the next exposure can be taken.
Uninterruptible Power Supply (UPS), disconnect the
system from mains during lightning storms.
The patient positioning lights are class II laser
products (21 CFR § 1040.10).
The patient positioning lights are class 1 laser
products (Standard IEC / EN 60825-1: 2007).
equipment must be installed and put into service
according to the specific EMC information provided in
the accompanying documents.
can affect the X-ray unit.
input, signal output or other connectors, shall comply
with relevant IEC standard (e.g. IEC 60950 for IT
equipment and the IEC 60601 series for medical
electrical
equipment).
combinations - systems - shall comply with the
standard IEC 60601-1-1, Safety requirements for
medical electrical systems. Equipment not complying
to IEC 60601 shall be kept outside the patient area
(more than 2m (79 in.) from the X-ray unit).
Any person who connects external equipment to
signal input, signal output or other connectors has
formed a system and is therefore responsible for the
system to comply with the requirements of IEC 60601-
1-1. If in doubt, contact your service technician or local
representative for help.
decrease in image quality.
exposure but the image does not appear in the
Planmeca Romexis program. The last ten images can
be manually imported into Romexis.
ray unit.
Planmeca ProMax 3D s / 3D 5
SAFETY PRECAUTIONS