Planmeca ProMax 3D S User Manual page 48

Dental x-ray unit, 3d imaging
Hide thumbs Also See for ProMax 3D S:
Table of Contents

Advertisement

INFORMATION DISPLAYS
Test exposure (i460)
Beam check (i470)
Network settings (i480)
44 Planmeca ProMax 3D s / 3D
During the process the words Exposure and Wait will
appear on the display. You can press and hold down the
exposure button for the whole duration of the process or
remove your thumb from the exposure button when the
word Wait appears on the display.
The seasoning process will take up to about three
minutes. After a successful process the message OK is
shown on the display.
NOTE Contact your service technician for help if error
message E332 (Arcing across X-ray tube) reappears
after a successful seasoning process.
Select Test exposure (i460) on the Special functions
(i400) display. The display shown below appears.
This option allows you to take a test exposure. The Test
exposure (i460) display shows the exposure values which
are in use on the main display.
Touch the arrows to select the exposure values (kV, mA
and time) you wish to use. Then touch the Test field. The
word Ready will appear at the bottom of the display and
you can now press the exposure button to take a test
exposure. Note that the C-arm will not move when you
take a test exposure.
Refer to the Technical Manual for more information on
beam check.
Refer to the Technical Manual for more information on
network settings.
User's Manual (3D)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Promax 3d

Table of Contents