Code
H161
Temperature
H162
H163
H165
H166
H171
User related
messages
H172
H175
H176
H177
H178
H180
H181
H182
H183
H184
H185
H186
H187
H189
H192
H195
User's Manual (3D)
Explanation
The temperature of the tube head is too
high.
The temperature of the lift motor is too
high.
The temperature of the power supply
unit (PSU) is too high.
The temperature of the tube head is too
high for the selected exposure values.
The maximum tube head energy level
was exceeded.
The sensor is not attached properly to
the C-arm.
The sensor is not attached properly to
the cephalostat.
PC program selection is in conflict with
the selected X-ray unit program.
Safety area limit violation in tomography
mode.
Exposure is not possible with these
settings.
Exposure is not possible with these
settings.
DEC is not available.
The imaging process was cancelled in
Romexis.
Timeout in image data transmission.
The attached sensor is not suitable for
the selected program.
Remove the 3D sensor.
The 3D sensor is not attached properly. Attach and / or lock the sensor
No IP address defined for 3D sensor.
Problem during image data
transmission.
The control panel was touched during
exposure.
It is not possible to enable radiation or
PC communication (i410) when demo
licenses are switched on (i510).
The 3D Model programs must not be
used for patient imaging.
HELP MESSAGES
Comments
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the lift
motor to cool down.
Wait for a few minutes for the
power supply unit (PSU) to cool
down.
Wait for a few minutes for the
tube head to cool down.
Wait for a few minutes for the
tube head to cool down or use
lower exposure values.
Attach and / or lock the sensor
in position.
Attach and / or lock the sensor
in position.
Select another exposure mode
in Romexis.
Change the values for layer
thickness, position or angle.
Change the settings for target
area.
Change the settings.
Exposure was interrupted.
Contact your service technician
for help.
Change the sensor.
in position.
Exposure was interrupted.
Contact your service technician
for help.
Exposure was interrupted.
First switch off demo licenses
(i510), then enable radiation or
PC communication (i410).
Use the 3D Model programs for
taking exposures of impressions
or plaster casts only.