CONTROL PANEL
8.6.3
3D Large View 3 horizontal
160mm
Volume
diameter & height
cannot be changed
for stitch volume 2
(Volume Diameter
field not shown on
Planmeca ProMax 3D s)
28 Planmeca ProMax 3D s / 3D
If you have selected the 3D program Large View 3
horizontal, select three target areas as follows:
Select the first two target areas as described earlier in
Large View
section 8.6.2 "3D Large View 2 horizontal" on page 26.
3 horizontal
Then continue to select the target area for stitch volume 2
by touching the Stitch volume 2 field which will now be
shown on the display. The Select stitch target area 2
display will appear.
Now set the target area for stitch volume 2 by touching
the desired position on the jaw icon. The yellow circle
(stitch volume 2) will move to the selected position.
NOTE FOR PLANMECA PROMAX 3D:
In horizontal stitching the diameter of the primary
volume is always Full. The diameter of a stitched
volume has to be set to Full as well.
The height (Lower, Upper, Full) selected for the primary
volume determines the height for the stitched volumes,
i.e. the volume height cannot be changed on the Select
stitch target area 2 display.
You can return to the previous display by touching the
Stitch volume 1 field if you need to readjust the target
area for stitch volume 1.
Confirm the selections by touching OK.
Back to stitch volume 1
Stitch volume 2=
yellow
Stitch volume 1 =
green
Primary volume =
red