3D EXPOSURE
12.3.2 Adjusting volume center lights (front and side lights)
Side light (y coordinate)
Front light
(x coordinate)
Volume center
lights
y
Patient support table
58 Planmeca ProMax 3D s / 3D
The volume center lights cross at the selected target
position. The cross formed by the light beams indicates
the center of the image volume. With the patient
positioned in the unit, the volume center lights form red
lines on the front (front light) and on the left side (side
light) of the patient's head.
Image
volume
Rotate the C-arm 90 degrees clockwise by touching the
90° field at the bottom of the display. This will give you a
better view for checking the volume position.
Touch the 90° field again to rotate the C-arm back to the
original position.
The position of the image volume can be fine-adjusted as
follows.
x
•
Thumb wheel
(side light =
y coordinate)
The side light (y coordinate) can be adjusted by
moving the thumb wheel that is located on the
underside of the patient support table.
The selected position is shown in the Target field on
the main display.