8.6.4
3D Large View 2 vertical
80mm
Target
movement 3 cm
(Volume Diameter
field not shown on
Planmeca ProMax 3D s)
Target movement
8.7

Selecting target coordinates

User's Manual (3D)
80mm
Target
movement 6 cm
If you have selected the 3D program Large View 2
vertical, the Select target area display will appear.
On the Select target area display, specify the target area
as described in section 8.6.1 "3D Standard and 3D
Horizontal Pair" on page 24. The selected target area will
be marked with a red circle on the jaw icon.
Touch the Target movement field to select the vertical
movement.
The movement can be either 3 or 6 cm. If you select 3 cm,
the maximum total height of the two volumes is 11 cm
(adult setting). If you select 6 cm, the maximum total
height of the two volumes is 13 cm (adult setting).
Confirm the selections by touching OK.
Touch the Target field on the main display to fine-adjust
the position of the selected target area. A display appears
where you can adjust the x / y coordinates by touching the
left / right arrows. The selected coordinates will be shown
on the display.
Touching either of the jaw icons on the right side of the
main display will take you to the Select target area display
where you can change any settings for the target area.
See section 8.6 "Selecting target area" on page 24 for
more information.