Access the [Pick/Install Xdcr] menu. Check the [Transducer Size] menu cell. If necessary, enter the
[Xdcr Size] option list and pick the transducer under test.
Check the [Xdcr Mount Mode] menu cell. Adjust to match the simulator chart above.
Move the highlight down to [Install Complete?] by triggering the <Down Arrow>. Access the option
list. Move the cursor to [Install] by triggering the <Up or Down Arrow>. Trigger the ENTER sensor to
start the transducer install routine.
NOTE: Since Sizes A and C transducers are installed in Reflect Mode, you will see a pop-up
window that prompts you to: [Use Actual Zero]. You can ignore this by triggering the
<Down Arrow>.
Upon the completion of the [Install] procedure, the computer should report a sonic velocity within the
range of approximately 1350m/s to 1700 m/s (depending on the transducer size under test). Next,
check the [Diagnostic Data] menu. The [Valc %] item must be >35 for dependable operation. In addi-
tion, note the reading, then compare it to the [Vf max] item in the [Diagnostic Data/Site Setup Data]
menu. The value of the reading should be less that 2% of the published Vf max. Confirming these values
certifies that the entire system (computer, transducers, cables) is operating correctly. The investigation
should proceed to a review of all site conditions to locate the operating problem.

6.9.2 USING THE 996PSP PIPE SIMULATOR

Using the proper coupling compound, mount the transducers on the pipe simulator as shown below.
Slide each transducer until it presses against a pin-stop. Use the clamping screws to hold the transduc-
ers in place.
1012TB-1 & 2 Universal Transducer Test Block Chart (English)
Xdcr
Pipe OD
Pipe
Size
(in)
Mat'l
A
0.650
Steel*
B
1.150
Steel*
C
2.000
Steel*
D
3.500
Steel*
1012TB-1 & 2 Universal Transducer Test Block Chart (Metric)
Xdcr Pipe OD
Pipe
Size (mm)
Mat'l
A
16.5
Steel*
B
29.2
Steel*
C
50.8
Steel*
D
88.9
Steel*
Wall
Mount
Spacing
Thk (in)
Mode
0.100*
Reflect
0.100*
Direct
0.100*
Reflect
0.100*
Direct
Wall
Mount
Thk (mm)
Mode
2.54*
Reflect
2.54*
Direct
2.54*
Reflect
2.54*
Direct
6-17
CQO:1010PVXFM-3
Number
Offset
Index
Nominal
7
Minimum
4
Nominal
11
Minimum
Use Ltn
Spacing
Number
Offset
Index
Nominal
7
Minimum
4
Nominal
11
Minimum
Use Ltn