Section 6
MESSAGE
Memory Full!
Memory Corrupted!
Chan Not Setup
Clr Active Memory?
Clr Saved Data?
<EOT>
Empty Pipe-Press [ENT] Prompt to empty the pipe during the Actual MTY procedure. After emptying
Fill Pipe - Press [ENT
No Sites
Not Installed
Re-space Index
Invalid Setup
(use Direct Mount)
Low Signal-Press [ENT] During the Initial Makeup, the meter decides that the level of the receive signal
Response to an attempt to save site data, when data memory is full. Delete an
obsolete site or clear Datalogger memory to make room for the new data.
Memory read error occurred while accessing the active site data.
Response to an attempt to invoke an operation that requires a channel to be
enabled. Enable the channel [Channel Setup
Note that a channel cannot be enabled until its transducers are operating.
Response to restoring operation if a severe event (e.g., a violent power surge)
disrupts system operation.
Response to resetting and selecting [No] for Clr All Data?
Response to a request to output datalogger data to the printer or the Graphics
screen when no datalogger data exists. Set up the Datalogger.
pipe, press <ENT>.
Prompt to fill the pipe during the Actual MTY procedure. After fill-
ing the pipe, press [ENT].
Response while trying to recall/delete a site setup when no Sites
are stored.
Response to an attempt to access a menu function that is not included.
Upon measuring the liquid sonic velocity (Vs), the meter recommends re-spac-
ing the transducers to improve performance.
During the Initial Makeup, the system detects invalid transducer spacing, erro-
neous liquid/pipe parameters, or some other factor that prevents it from com-
pleting the Initial Makeup. This may be due to one of the following:
• An out-of-range data entry.
• An invalid condition; (e.g., overlapping transducers in Reflect Mode). If select-
ing Direct Mode does not resolve, review all site setup and transducer installa-
tion choices; particularly data entered for the pipe and liquid.
• In Reflect Mode, the flow computer detects that the pipe wall signal may
infringe upon the liquid signal. Use Direct Mode instead.
Press <ENT>, <UP Arrow>, <DOWN Arrow>, or <LEFT Arrow> to abort install
routine. Continue programming other site data in anticipation of resolving the
difficulty later. Call technical support for help if necessary.
is insufficient for proper operation. Some reasons for low signal are:
• Invoking [Install Completed?] on an empty pipe.
• Coupling compound insufficient, not applied, or evaporated.
• You need to condition the pipe at the mounting location.
• A large air bubble needs to be "flushed out."
• The Xdcr cables are defective or not connected to the correct channel.
• The Set Empty routine performed when pipe was NOT actu-ally empty.
If you locate and correct the improper condition immediately, press <ENT> to
resume the installation procedure. Otherwise, press the <LEFT Arrow> to abort
the installation and conduct a thorough investigation.
DESCRIPTION
6-12
CQO:1010PVXFM-3