Section 5
5.10.2 THE TABLE SETUPS MENU
The [Table Setups] menu cell allows you to pre-condition your pipe table and transducer types. The edits
made in [Table Setups] become the default settings for creating a new site. Transducers "marked" in the
[Transducer Types] menu will be preferentially selected when the meter recommends transducers dur-
ing the automatic Pick/Install Transducer routine.
To access the [Table Setups] menu press
Pipe Table
The factory-programmed pipe table describes over sixty standard English and metric pipes. The table
can save you programming steps by loading all required data at once from option list selection. The Pipe
Table configuration is [PIPE CLASS
English Pipe classifications. Selecting a class (e.g., ASA Carbon Steel) conditions the Pipe Name option
list to all the stored pipes within that class (see Pipe Data Menu Structure on next page). Selecting a pipe
by its name (e.g., 2SS10) loads a description of that pipe in the remaining menu cells. Note that liner
data is not provided. If your pipe is lined, then you will be required to enter the liner material and its
thickness manually.
To access the [Pipe Table] menu press
Table Setup
PIPE NAME]. Pipe Class presents a list of standard metric and
5-66
CQO:1010PVXFM-3
Pipe Table
Transducer Type