Type digit 9 and then press <ENT>. Next, move the highlight to [Wall Thickness]. Press the <Right
Arrow>, type 0.29, and then press <ENT>. The [Pipe Data] screen will show these changes as seen
below.
Press the <Left Arrow> to go back to the [Clamp-On] menu. Move the highlight up to [Channel
Setup]. Press the <Right Arrow> and move the highlight to [Create/name Site].
Create a site named: SITE1 (as detailed in paragraph 2.8.1).
The above were a only a few examples of the Installation Menu functions. Refer to Section 5 for
more detailed step-by-step instructions.
Siemens
Site
Enter pipe Outer Diameter manually
Pick Pipe Class
Select Pipe Size
Pipe OD <in>
Pipe Material
Wall Thickness
Liner Material
Liner Thickness
Pipe Data
Siemens
Site
Identify Liner material for Lined Pipes
Pick Pipe Class
Select Pipe Size
Pipe OD <in>
Pipe Material
Wall Thickness
Liner Material
Liner Thickness
Pipe Data
2-32
[1]
SITE1
Manual Entry
N/A
=0.500
Steel
0.100
None
0.000
[1]
SITE1
Manual Entry
N/A
9.000
Steel
0.290
None
0.000