Force Transmit Procedure - Siemens SITRANS FUH1010PVDX Operating Instructions Manual

Ultrasonic flowmeters ip65 nema 7 compact precision volume 7me3601-3
Table of Contents

Advertisement

Section 5
Press <Down Arrow> and scroll to the [Install Complete?] menu cell and then press <ENT> (unless
otherwise directed to do so by the Technical Service Department).
Press <ENT>. The [Install Complete?] menu cell will indicate [Yes] after the FUH1010PVX is success-
fully in-stalled.
To complete the Install process disregard the following paragraphs explaining the Force
Transmit and Force Frequency diagnostic software routines. Proceed to the Install pro-
cess completion steps immediately following the Force Transmit procedures.
NOTE: If the transducers have been installed successfully but the Estimated Vs (sonic velocity)
has been changed, the Pick/Install Xdcr menu cell [Install Complete?] option list will also
display the [Transfer Install] function selection. The Transfer Install function allows the
transducers to be optimally positioned for a different fluid, without the need for a new
Initial Makeup procedure (see the Water Calibration addendum in this manual for de-
tails).

Force Transmit Procedure

This diagnostic software routine allows the user to "force" a transmitting condition that can be used to
search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The routine
forces the flowmeter to generate constant transmit bursts while reporting current receive signal strength
for the user. To initiate the Force Transmit function, refer to the example that follows:
Siemens
Dual Path
Key [Install] after mounting transducers
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Complete?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
Siemens
Dual Path
Key [Install] after mounting transducers
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Complete?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
5-27
SITE1
1011 Universal
B3
Direct
Minimum
4
Track 1012TP
0.217
No
Channel Not Setup
Channel Not Setup
SITE1
1011 Universal
B3
Direct
Minimum
4
Track 1012TP
0.217
Yes
Channel Not Setup
Channel Not Setup
CQO:1010PVXFM-3