Siemens SITRANS FUH1010PVDX Operating Instructions Manual page 69

Ultrasonic flowmeters ip65 nema 7 compact precision volume 7me3601-3
Table of Contents

Advertisement

Section 2
Press the <Right Arrow> to move into the [Channel Setup] menu as shown below. Note that the
highlight is on the first menu cell: [Recall Site].
Use the <Down Arrow> to move the highlight to [Create/Name Site]. Create a site named: SITE1 (as
detailed in paragraph 2.8.1).
2.9.2 HOW TO ENTER PIPE DATA
After creating the site, the next step is to enter the Pipe Data. The Pipe Data menu allows you to either
enter the pipe data manually or select a preset pipe configuration from an option list. This section ex-
plains how to enter the data manually. The two mandatory entries to be entered in this menu are the
pipe's Outer Diameter (OD) and its Wall Thickness. Here we will insert a pipe outer diameter of 9.00"
and a pipe wall thickness of 0.29" as an example.
Press the <Left Arrow> to leave [Channel Setup]. Press the <Down Arrow> and then the <Right
Arrow> to enter the [Pipe Data] menu.
Move the highlight to [Pipe OD <in>] and then press the <Right Arrow> and then the <Right Arrow>.
Note that an equal sign appears before the current parameter.
Siemens
Create-Name-Recall-Enable & Delete Site
Channel Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
DataSpan/Set/Cal
Datalogger Setup
I/O Data Control
Diagnostic Data
Siemens
Scroll saved Site Setup list and Enter
Recall Site
Channel Enable
Create/Name Site
Site Security
Delete Site
Save/Rename Site
Channel Setup
2-31
Site
[1]
Site
[1]
No Sites
No
Off
No Sites
CQO:1010PVXFM-3