Basic Function Commands
List/Sweep Subsystem ([:SOURce])
This command sets a list or step sweep point as the current sweep point
controlling the frequency and power output.
If list or step mode is controlling frequency or power, or both, then the indexed
point in the respective list(s) will be used.
Entering a value with this command will have no effect, unless MANual is the
selected mode. Refer to the :LIST:MODE command for setting the proper
mode.
If the point selected is beyond the length of the longest enabled list, then the
point will be set to the maximum possible point, and an error will be generated.

:LIST:MODE

[:SOURce]:LIST:MODE AUTO|MANual
[:SOURce]:LIST:MODE?
This command sets the operating mode for the current list or step sweep.

:LIST:OPTions

[:SOURce]:LIST:OPTions <val>,<val>
[:SOURce]:LIST:OPTions?
[:SOURce]:LIST:OPTions:POINts?
This command enables specific options during a list sweep. The command
adds the capability to suppress FM, fM, and AM on any list sweep point.
Additionally, frequency, power, and/or a waveform transition can be
suppressed resulting in no synthesizer, no output, or no waveform playing
interruption during a transition.
86
Range
List Sweep: 1 to 3,201
Key Entry
Manual Point
Supported
All Models
AUTO
This choice enables the selected sweep type to perform
a sweep of all points.
MANual
This choice enables you to select a single sweep point.
The selected point controls the frequency and/or
amplitude according to the sweep type. Refer to the
:LIST:MANual command for selecting a sweep point.
*RST
AUTO
Key Entry
Manual Mode Off On
Supported
All Models
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference
Step Sweep: 2 to 65535