Basic Function Commands
List/Sweep Subsystem ([:SOURce])
:LIST:DWELl:TYPE
[:SOURce]:LIST:DWEL1:TYPE LIST|STEP
[:SOURce]:LIST:DWEL1:TYPE?
This command toggles the dwell time for the list sweep points between the
values defined in the list sweep and the value for the step sweep.
:LIST:FREQuency
[:SOURce]:LIST:FREQuency <value>,<value>
[:SOURce]:LIST:FREQuency?
This command sets the frequency values for the current list sweep points.
The maximum number of list sweep points is 3,201.
The variable <value> is expressed in units of hertz (Hz).
The setting enabled by this command is not affected by signal generator
power–on, preset, or *RST.
:LIST:FREQuency:POINts
[:SOURce]:LIST:FREQuency:POINts?
This command queries the current list sweep file for the number of frequency
points.
:LIST:MANual
[:SOURce]:LIST:MANual <value>|UP|DOWN
[:SOURce]:LIST:MANual?
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference
Supported
All Models
LIST
This choice selects the dwell times from the list sweep.
Refer to the :LIST:DWELl command for setting the list
dwell points.
STEP
This choice selects the dwell time from the step sweep.
Refer to the :SWEep:DWELl command for setting the
step dwell.
*RST
LIST
Key Entry
Dwell Type List Step
Supported
All Models
Range
The range is model/option dependent. Refer to the
instrument's Data Sheet.
Supported
All Models
Supported
All Models
85