Analog Modulation Commands
Frequency Modulation Subsystem–Option UNT ([:SOURce])
:FM[1]|2:INTernal:FUNCtion[1]|2:SHAPe:RAMP
[:SOURce]:FM[1]|2:INTernal:FUNCtion[1]|2:SHAPe:RAMP
POSitive|NEGative
[:SOURce]:FM[1]|2:INTernal:FUNCtion[1]|2:SHAPe:RAMP?
This command sets the ramp direction when
:FM[1]|2:INTernal:FUNCtion[1]|2:SHAPe is set to RAMP.
Key Entry
:FM[1]|2:INTernal:NOISe[1]|2:TYPe
[:SOURce]:FM[1]|2:INTernal:NOISe[1]|2:TYPe UNIForm|GAUSsian
[:SOURce]:FM[1]|2:INTernal:NOISe[1]|2:TYPe?
This command sets the noise type when
set to NOISe[1] or NOISe2.
Key Entry
:FM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STARt
[:SOURce]:FM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STARt
<val><units>
[:SOURce]:FM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STARt?
This command sets the start frequency for the swept function generator.
:FM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STOP
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference
Remarks
The waveform selection is only valid when INT is the
source selection. Refer to
270
for source type selection.
SINE is the only shape available without Option 303.
Supported
All Models with Options UNT and 303
*RST
POS
Pos Ramp
Supported
All Models with Options UNT and 303
*RST
POS
Noise Gen 1
Supported
All Models with Options UNT and 303
*RST
400.0 Hz
Range
100 mHz to 6.250 MHz
Key Entry
FM Start Freq
Supported
All Models with Options UNT and 303
":FM[1]|2:SOURce" on page
Neg Ramp
":FM[1]|2:SOURce" on page 270
Noise Gen 2
is
265
Need help?
Do you have a question about the N5171B EXG and is the answer not in the manual?