Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:STEP:STARt
[:SOURce]:CORRection:FLATness:STEP:STARt <freq><unit>
|MAXimum|MINimum|DEFault|
[:SOURce]:CORRection:FLATness:STEP:STARt? [MAXimum|MINimum]
This command sets the start frequency for the user flatness calibration step
array. See also, :FLATness:STEP:POINts and :FLATness:STEP:STOP commands.
:FLATness:STEP:STOP
[:SOURce]:CORRection:FLATness:STEP:STOP <freq><unit>
|MAXimum|MINimum|DEFault|
[:SOURce]:CORRection:FLATness:STEP:STOP? [MAXimum|MINimum]
This command sets the stop frequency for the user flatness calibration step
array. See also, :FLATness:STEP:POINts and :FLATness:STEP:STARt
commands.
:FLATness:STORe
[:SOURce]:CORRection:FLATness:STORe "<file name>"
This command stores the current user–flatness correction data to a file named
by the :CORRection:FLATness:STORe command. The directory path is
implied in the command and need not be specified in the "<file name>"
variable.
52
Range
2 to 3,201
Key Entry
# Points
Supported
All Models
*RST
The preset value is model/option dependent. Refer to
the instrument's Data Sheet.
Range
The range is model/option dependent. Refer to the
instrument's Data Sheet.
Key Entry
Freq Start
Supported
All Models
*RST
The preset value is model/option dependent. Refer to
the instrument's Data Sheet.
Range
The range is model/option dependent. Refer to the
instrument's Data Sheet.
Key Entry
Freq Stop
Supported
All Models
Key Entry
Store To File
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference