System Commands
Trigger Subsystem

:TRIGger[:SEQuence]:TIMer

:TRIGger[:SEQuence]:TIMer <period>
:TRIGger[:SEQuence]:TIMer?
This command sets the period of the timer trigger.

:TRIGger[:SEQuence][:IMMediate]

:TRIGger[:SEQuence][:IMMediate]
This event command causes an armed List or Step sweep to immediately start
without the selected trigger occurring.

:TSWeep

[:SOURce]:TSWeep
This command aborts the current sweep, then either arms or arms and starts a
single list, depending on the trigger type.
The command performs the following:
— arms a single sweep when BUS, EXTernal, or Trigger KEY is the trigger
— arms and starts a single sweep when IMMediate is the trigger source
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference
Supported
All Models
*RST
1 ms
Range
.5 ms to 1000 seconds
Key Entry
Trig Timer Period
Supported
All Models
Supported
All Models
source selection
selection
Key Entry
Single Sweep
245