[:Connectors]:Bbtrigger[1]|2[:Output]; [:Connectors]:Ptrig[:Output] - Keysight N5171B EXG Manual

X-series signal generators
Hide thumbs Also See for N5171B EXG:
Table of Contents

Advertisement

System Commands
Route Subsystem (:ROUTe)
:ROUTe:LINE:PTRigger[1]|2:BNC:SOURce
BBTRigger[1]|BBTRigger2|EVENt[1]|PTRigger|NONE
:ROUTe:LINE:PTRigger[1]|2:BNC:SOURce?
This command selects a BNC connector to use as an input for the Pattern
Trigger In source.

[:CONNectors]:BBTRigger[1]|2[:OUTPut]

:ROUTe[:CONNectors]:BBTRigger[1]|2[:OUTPut]
M1|M2|M3|M4|AUX29|NONE
:ROUTe[:CONNectors]:BBTRigger[1]|2[:OUTPut]?
This command selects a marker (M1–M4) signal to be routed to the specified
rear panel BB TRIG 1 or BB TRIG 2 connector. AUX29 selects Aux I/O Pin 29.
NONE indicates that the BNC connector is, or can be, an input.

[:CONNectors]:PTRig[:OUTPut]

:ROUTe[:CONNectors]:PTRig[:OUTPut] M1|M2|M3|M4|AUX29|NONE
:ROUTe[:CONNectors]:PTRig[:OUTPut]?
This command selects a marker (M1–M4) signal to be routed to the rear panel
PAT TRIG connector. AUX29 selects Aux I/O Pin 29. NONE indicates that the
BNC connector is, or can be, an input.
194
This choice sets the BB TRIG 1 connector as the Pattern
BBTRigger1
Trigger In source.
This choice sets the BB TRIG 2 connector as the Pattern
BBTRigger2
Trigger In source.
This choice sets the EVENT 1 connector as the Pattern
EVENT1
Trigger In source.
This choice sets the PAT TRIG connector as the Pattern
PTRigger
Trigger In source.
This choice selects no Pattern Trigger In source.
NONE
*RST
PTR
Key Entry
Patt Trig Source
Supported
N5172B with Option 653 or 655, N5182B with Option
656 or 657
*RST
BBTRigger1 - M2
BBTRigger2 - None
Key Entry
Route to BB TRIG 1, 2 BNC
Supported
N5172B with Option 653 or 655, N5182B with Option
656 or 657
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference