System Commands
Calibration Subsystem (:CALibration)

:IQ:TYPE

:CALibration:IQ:TYPE DC|USER|FULL
:CALibration:IQ:TYPE?
This command sets the IQ calibration type.

:IQ:[:USER]

:CALibration:IQ[:USER]
This command performs a IQ calibration according to the IQ calibration type.
For information on selecting the type of IQ calibration, refer to
page
This calibration should be run when the ambient temperature has varied by at
least ±5 degrees celsius from the ambient temperature at which the previous
calibration was run.
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference
Key Entry
Stop Frequency
Remarks
The setting enabled by this command is not affected by
signal generator power–on, preset, or *RST.
Supported
N5172B with Option 653 or 655, N5182B with Option
656 or 657
Key Entry
Calibration Type DC User Full
Remarks
The setting enabled by this command is not affected by
signal generator power–on, preset, or *RST.
Supported
N5172B with Option 653 or 655, N5182B with Option
656 or 657
135.
Key Entry
Execute Cal
":IQ:TYPE" on
135