Analog Modulation Commands
Pulse Modulation Subsystem–Option UNU and UNW and Option 320 ([:SOURce])
:PULM:INTernal:TRAin:REPetition:POINts?
[:SOURce]:PULM:INTernal:TRAin:REPetition:POINts?
This query returns the count of elements in the list of repetitions.
:PULM:INTernal:TRAin:TRIGger
[:SOURce]:PULM:INTernal:TRAin:TRIGger FRUN|TRIGgered|GATEd
[:SOURce]:PULM:INTernal:TRAin:TRIGger?
This command sets the triggering mode for the Pulse Train feature.
:PULM:INTernal:TRAin:TRIGger:IMMediate
[:SOURce]:PULM:INTernal:TRAin:TRIGger:IMMediate
[:SOURce]:PULM:INTernal:TRAin:TRIGger:IMMediate?
This command will cause the pulse train to run once. If the pulse train is
already running or off, then this SCPI command has no effect.
:PULM:INTernal:VIDeo:POLarity
Keysight EXG and MXG X-Series Signal Generators SCPI Command Reference
Supported
All with Options UNW and 320
Supported
All with Options UNW and 320
FRUN
Free Run triggering continuously plays the pulse train.
TRIGgered
Trigger runs the pulse train (after waiting the Pulse
Delay) each time an external trigger is supplied (edge
triggered) to the PULSE BNC, the "Trigger Immediately"
softkey is pressed, or the
[:SOURce]:PULM:INTernal:TRAin:TRIGger:IMMediate
SCPI command is sent (when the pulse train playback is
idle). Triggers received during playback are lost.
GATEd
Gated triggering runs the pulse train while an external
trigger is supplied (level triggered) to the PULSE BNC.
The state of the GATEd trigger is detected only when
the playback is transitioning to or in idle. This means
that, once started, playback is always completed, even
if the GATE trigger changes to the inactive state.
*RST
TRIGgered
Key Entry
Trigger Mode
Supported
All with Options UNW and 320
Key Entry
Trigger Immed iately
Supported
All with Option UNW