General Fault Conditions - Abbott CELL-DYN 3000 Operator's Manual

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Troubleshooting
10-62
Search
Book TOC
Table 10.2
General Fault Conditions (Continued)
Message acknowledgment time out/BL
– – – – – Probable Cause – – – – –
1.
Communication between
the Analyzer and the Data
Station did not occur when
expected.
__________________________________________________________
Message reception time out/BL
– – – – – Probable Cause – – – –
1.
Communication between
the Analyzer and the Data
Station did not occur when
expected.
__________________________________________________________
– – – – – Probable Cause – – – –
1.
An illegal software
operation was requested
by the Analyzer.
Go Back
– – – – Corrective Action – – – –
1. Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second
MENU
recurs, call for technical
assistance.
– – – – Corrective Action – – – –
1. Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second
MENU
recurs, call for technical
assistance.
Run time error/BL
– – – – Corrective Action – – – –
1. Document what was
happening when the
message was displayed and
call for technical assistance.
2. Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second
MENU
processing if the fault does
not recur.
CELL-DYN® 3000 System Operator's Manual
Chapter 10
DIAGNOSTICS
Screen. If the fault
DIAGNOSTICS
Screen. If the fault
DIAGNOSTICS
Screen. Resume
9140240E — May 1995