Chapter 5

Procedure Customize QC Log Printout

STANDARD
SELECTION
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Search
Book TOC
CNT1 and CNT2 —
The [STANDARD SELECTION] key is used to automatically arrange
the parameters in a predetermined print group.
1. Select a file from the
cursor to the desired file.
2. Press [CUSTOMIZE PRINTOUT] to display the
QC PRINTOUT Screen
3. Use the Arrow keys on the keyboard to move the cursor to the
desired parameter in the list under the printout group.
4. Press [SELECT PARAMETER]. The selected parameter is
highlighted and the cursor moves to the first position in the group.
NOTE: The key label changes to [PLACE PARAMETER]
and a [CANCEL SELECTION] key is displayed.
5. If necessary, use the Arrow keys on the keyboard to move the
cursor to the desired location and press [PLACE PARAMETER].
NOTE: When the [PLACE PARAMETER] key is pressed, the
selected parameter is displayed in the position indicated by the
cursor and the cursor then advances to the next parameter in the
list under the printout group.
6. Repeat steps 3–5 until all entries have been made.
7. If desired, press the Print Screen key on the keyboard to obtain a
printout of the configuration.
8. Press [RETURN] to return to the
9. Repeat this procedure to customize the printout for other QC logs.
When the [STANDARD SELECTION] key is pressed, the parameters
are automatically arranged in the predetermined print group shown in
Figure
5.14.
Operating Instructions
Go Back
CNT1 is the RBC/PLT Count Time. CNT2 is
the RBC/PLT Count Time for an extended
count.
QC SETUP MENU Screen
for the selected file.
QC SETUP MENU Screen.