Chapter 6
MAIN
ENTER
FACTOR
RESTORE
FACTORS
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Search
Book TOC
The [MAIN] key is used to return to the
The [ENTER FACTOR] key displays the
FACTOR Screen
Factors (See Figure 6.3). Calibration factors may be changed on this
screen by moving the cursor to the desired position, typing the new factor
and pressing the Enter key on the keyboard. The following soft key labels
are displayed:
RESTORE FACTORS
RESET ALL TO 1.00
RETURN
The [RESTORE FACTORS] key is used to restore the previous
calibration factors. This key is only active immediately after factors have
been changed.
Parameter
(Factor Range)
WBC
(0.70..1.30)
RBC
(0.80..1.20)
HGB
(0.70..1.30)
MCV
(0.70..1.30)
PLT
(0.70..1.30)
MPV
(0.70..1.30)
RESTORE
RESET ALL
FACTORS
TO 1.00
Figure 6.3:
Go Back
showing the current Open and Closed Calibration
ENTER WHOLE BLOOD FACTOR
Ready
Open Sampler
Factor
1.008
1.071
1.124
0.985
1.028
1.000
Enter Whole Blood Factor Screen
Calibration
MAIN MENU
Screen.
ENTER WHOLE BLOOD
May 11 1992
10:34
Operator ID
734
Sequence #
4100
Closed Sampler
Factor
1.056
1.060
1.136
0.985
0.939
1.000
RETURN
6-15