Calibration For Individual Parameters; Completing Mode To Mode Calibration - Abbott CELL-DYN 3000 Operator's Manual

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Calibration

Calibration for Individual Parameters

Completing Mode to Mode Calibration

6-78
Search
Book TOC
5. Type any comments in the <COMMENTS> field. Press the Enter
key on the keyboard to save the entry and advance the cursor.
Comments may be added to the Calibration Log only after a
calibration factor is changed or reentered.
6. If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
If only certain individual parameters require calibration, continue with
this section:
1. Press [QUIT AUTO-CAL] followed by [CONFIRM QUIT] to
exit Auto-Cal.
2. Press [RETURN] twice to return to the
Screen.
3. Press [ENTER FACTOR].
The
ENTER WHOLE BLOOD FACTOR Screen
4. Type any mean factors computed by Auto-Cal for those parameters
that require calibration in the appropriate fields. (Refer to the
printout made in
section of this procedure.) Move the cursor to the desired position,
type the factor and press the Enter key on the keyboard to save the
factor and advance the cursor.
5. Press [RETURN]
6. When the
CALIBRATION LOG Screen
parameters that required calibration in the <COMMENTS> line.
Press the Enter key on the keyboard to save the entry and advance
the cursor.
NOTE: Manually entered factors are followed by the letter E in
parentheses to indicate "Enter Factor".
7. If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
1. Press [RETURN] to return to the
Screen.
2. Press [MAIN] to exit the
Go Back
step 2
of the
"Calibration Factor
CALIBRATION MENU
CALIBRATION
CELL-DYN® 3000 System Operator's Manual
Chapter 6
CALIBRATION MENU
is displayed.
Calculation"
is displayed, indicate the
MENU.