Abbott CELL-DYN 3000 Operator's Manual page 324

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Chapter 6
Instrument:__________________________________ Date:______________________________________
Operator:______________________________________
________________________________________________________________________________________
Factor
%Diff*
WBC
RBC
HGB
MCV
PLT
* Delete the sign of the Factor % Diff before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Search
CELL-DYN 3000
Mode to Mode Calibration Worksheet
Mode to Mode Calibration Criteria
Validation Range
Cal Not Required
<1.75%
<1.25%
<1.25%
<1.25%
<3.50%
Book TOC
Go Back
Cal Range
Cal Required
>1.75% but
<
10%
<
>1.25% but
10%
>1.25% but
<
10%
<
>1.25% but
10%
<
>3.50% but
20%
Calibration
Cal Limit
Cal?
Do Not Cal**
Y/N
>10%
>10%
>10%
>10%
>20%
6-69