Abbott CELL-DYN 3000 Operator's Manual page 345

Hide thumbs Also See for CELL-DYN 3000:
Table of Contents

Advertisement

Calibration
6-90
Search
Book TOC
• Analyzer Module Set Points (e.g., gains, dil factors [internal
calibration factors], thresholds, pressure/vacuum settings)
• RBC Editing Ratio
To back up the calibration factors, proceed as follows:
1. Turn the Data Station power switch OFF.
2. Turn the Analyzer power switch OFF.
3. Obtain the Set-Up Disk from the diskette box located behind the
upper front cover of the Analyzer.
4. Insert the Set-Up Disk in the Data Station disk drive.
5. Turn the Data Station power switch ON.
6. The DATA STATION Screen displays the following:
THIS DISK CAN BE USED TO COPY SET UP
INFORMATION TO OR FROM THE DATA STATION HARD
DISK. THE FOLLOWING COMMANDS ARE AVAILABLE:
GET [ENTER] - COPIES SET-UP INFORMATION FROM
THE DATA STATION HARD DISK ONTO THE SET-UP
DISK
PUT [ENTER] - COPIES SET-UP INFORMATION FROM
THE SET-UP DISK ONTO THE DATA STATION HARD
DISK
7. Type GET.
8. Press the Enter key on the keyboard. The Set-Up information is
copied from the Data Station Hard Disk onto the Set-Up Disk.
Previous Set-Up information that was stored on the disk is
overwritten.
9. When the copy process is complete, the "a:>" prompt is displayed
on the screen.
10. Remove the Set-Up Disk from the disk drive and return it to the
diskette box.
11. Turn the Data Station power switch OFF.
12. Wait five seconds and then turn the Analyzer power switch ON.
13. Turn the Data Station power switch ON.
14. When the Data Station Status Box indicates INITIALIZED, press
the [RUN] soft key to Prime.
Go Back
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Chapter 6