Data Station Run Screen Soft Keys

CLEAR
ORIFICE
CLEAR
FAULT
5-58
Search
Book TOC
A brief description of the function of each key is given in this section.
Instructions for running samples and using the Work List are given in the
Sample Analysis sections of this chapter.
The [CLEAR ORIFICE] key is used to initiate a special cleaning
sequence that flushes the RBC/PLT aperture to remove obstructions. The
sequence takes approximately 35 seconds. When the [CLEAR
ORIFICE] key is pressed, the message CLEANING RBC ORIFICE is
displayed in the Status Box.
The [CLEAR ORIFICE] key label changes to [CLEAR FAULT]
whenever a system fault occurs (e.g., Diluent Empty). This key is used to
clear the fault message and return the Analyzer to the "Ready" status
after corrective action has been taken.
NOTE: A message describing the fault appears in the bulletin
line. A list of fault conditions and corrective action is given in
Chapter 10,
Work List OFF
Bar Code On
#
4-DIGIT
SPECIMEN
BAR CODE
ID
793
1793
794
1794
795
1795
796
1796
797
1797
798
1798
799
1799
800
1800
WORK LIST
BAR CODE
INSERT/
ON
OFF
DELETE
Figure 5.27:
Go Back
Troubleshooting, of this manual.
WORK LIST
Ready
SPECIMEN
NAME
DELETE
PURGE
ALL
COMPLETED
Work List Screen
Dec 18 1992
09:48
Operator ID
734
Sequence #
0630
LIMIT PARAM STATUS RACK-TUBE
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
WORK LIST
PRINT
SET UP
WORK LIST
CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Chapter 5
RETURN