Chapter 9

Closed Sampler Tube Retainer Adjustment

CELL-DYN® 3000 System Operator's Manual
9140240E — May 1995
Search
Book TOC
It is necessary to adjust the tube retainer on the CELL-DYN 3000CS to
accommodate different sized Vacutainer® tubes. The closed sampler
module is depicted in Figure 9.16.
Tube
Retainer
Cap Piercer
Well
Figure 9.16:
Closed Sampler Module
Procedure
1. Squeeze the release levers on the sides of the tube retainer between
the thumb and forefinger to loosen it and slide the clamp up.
2. Insert the Vacutainer tube cap-down into the cap piercer well.
3. Slide the clamp down to hold the tube snugly in place.
4. Release the levers when the tube retainer has been properly
positioned.
5. Check the height adjustment with several Vacutainer tubes to be
certain that it is correct. The tube should fit snugly into the tube
retainer.
Go Back

Special Procedures

Maintenance
Release Levers
(One on each side)
Touch Plate
9-43