2. The cursor is in the <LOT NUMBER> entry field. Type the lot
number and press the Enter key on the keyboard to save the entry
and advance the cursor to the <EXPIRATION DATE> entry field.
3. This is the same format selected on the
Screen. Separate the digits with a slash (/) or a period (.).
4. Press the Enter key on the keyboard to save the entry and advance
the cursor to the <WESTGARD RULE SELECTION> entry
fields.
5. Use the Arrow keys on the keyboard to position the cursor at the
desired Westgard Rule.
6. Press [TOGGLE ON/OFF] to enable the rule and advance the
cursor.
7. Repeat steps 5 and 6 until all desired rule selections have been
made.
8. If desired, press [PRINT] to obtain a printout of the entries.
9. Press [RETURN] to return to the
Procedure Replicate ID Entry
1. If necessary, from the
[REPLICATE ID] to display the <REPLICATE ID> entry field.
2. The cursor is in the <REPLICATE ID> entry field. Type the
Replicate ID number (up to 12 alphanumeric characters may be
entered) and press the Enter key on the keyboard to save the entry
and advance the cursor to the <WESTGARD RULE
SELECTION> entry fields.
3. Select the Westgard Rules as directed in the
procedure.